Instlib API
|
|
-----------
|
|
|
|
Although instlib is writen in the C language, it uses an
|
|
object oriented style of API.
|
|
|
|
The instlib API is declared in the inst.h file, with suplimental
|
|
definitions in other .h files such as insttypes.h, disptechs.h,
|
|
icoms.h, xspect.h etc.
|
|
|
|
1) Instrument discovery
|
|
--------------------
|
|
|
|
A list of potential instruments is returned by a call to new_icompaths().
|
|
|
|
The diagnostic log argument should be set to the default global log g_log, if
|
|
it is not being overridden to use a different log.
|
|
|
|
The list of instruments can then be itterated through, to select which
|
|
instrument to use:
|
|
|
|
icompaths *icmps;
|
|
|
|
icmps = new_icompaths(g_log);
|
|
|
|
if (icmps != NULL && icmps->paths != NULL) {
|
|
int i;
|
|
for (i = 0; ;i++) {
|
|
if (icmps->paths[i] == NULL)
|
|
break;
|
|
printf(" %d = '%s'\n",i+1,icmps->paths[i]->name);
|
|
}
|
|
}
|
|
|
|
The chosen instrument device path can be retrieved using the "comport"
|
|
index 1..N, which corresponds to the index in the icompaths list (see above):
|
|
|
|
icompath *ipath;
|
|
|
|
if ((ipath = icmps->get_path(icmps, comport)) == NULL)
|
|
... error ...
|
|
|
|
|
|
2) User interface callback function
|
|
--------------------------------
|
|
|
|
There is the option of setting a user interface callback function,
|
|
either at the creation of the instrument object, or afterwards.
|
|
|
|
The callback is used for one of four different purposes:
|
|
|
|
To poll for a user abort during communication setup.
|
|
To signal that the instrument measurement has been triggered.
|
|
To poll for an abort while waiting to trigger.
|
|
To poll for a user abort during measurement.
|
|
|
|
The callback function will have the purpose paramater set appropriately.
|
|
|
|
For inst_negcoms, the return value of inst_user_abort
|
|
will abort the communication negotiation.
|
|
|
|
For inst_armed return value should be one of:
|
|
inst_ok to do nothing, inst_user_abort to abort the measurement,
|
|
or inst_user_trig to trigger the measurement.
|
|
|
|
For inst_triggered, the return value of the callback is ignored.
|
|
|
|
For inst_measuring the return value should be one of:
|
|
inst_ok to do nothing, or inst_user_abort to abort the measurement.
|
|
|
|
typedef struct _uicontext {
|
|
... your context here ...
|
|
} uicontext;
|
|
|
|
uicontext cb_cntx;
|
|
|
|
inst_code ui_cb_func(void *cntx, inst_ui_purp purp) {
|
|
uicontext *p = (uicontext *)cntx;
|
|
|
|
/* Negotiating communications with instrument. */
|
|
if (purp == inst_triggered) {
|
|
|
|
// Can abort by returning inst_user_abort;
|
|
|
|
return inst_ok;
|
|
|
|
/* Instrument is armed and ready for measurement */
|
|
} else if (purp == inst_armed) {
|
|
|
|
// Can trigger by returning inst_user_trig;
|
|
// Can abort by returning inst_user_abort;
|
|
|
|
return inst_ok; // Wait some more
|
|
|
|
/* Instrument measurement has been triggerted
|
|
} else if (purp == inst_triggered) {
|
|
|
|
... show in user interface that the instrument measurement
|
|
has triggered ...
|
|
|
|
return inst_ok;
|
|
|
|
/* Instrument is busy measuring */
|
|
} else if (purp == inst_measuring) {
|
|
|
|
// Can abort by returning inst_user_abort;
|
|
|
|
return inst_ok;
|
|
}
|
|
return inst_ok;
|
|
}
|
|
|
|
|
|
3) Opening instrument
|
|
------------------
|
|
|
|
Given the instrument device path ipath (see above), first create an instrument
|
|
object. A user interface callback function can be provided (see above).
|
|
|
|
inst *it;
|
|
|
|
if ((it = new_inst(ipath, 0, g_log, ui_cb_func, cb_cntx)) == NULL)
|
|
... error ...
|
|
|
|
and establish communications:
|
|
|
|
baud_rate br = baud_38400; /* Target baud rate - ignored for i1pro */
|
|
flow_control fc = fc_nc; /* flow control - ignored for i1pro */
|
|
|
|
if ((rv = it->init_coms(it, br, fc, 15.0)) != inst_ok)
|
|
... error ...
|
|
|
|
and initialise the instrument:
|
|
|
|
if ((rv = it->init_inst(it)) != inst_ok)
|
|
... error ...
|
|
|
|
|
|
4) Getting instrument capabilities and configuration
|
|
-------------------------------------------------
|
|
|
|
Because instlib handles many different types of instruments,
|
|
it is very useful to be able to determine what capabililities
|
|
an instrument has, to determine what sort of measurements it is
|
|
capable of making, and how to best configure and operate it.
|
|
|
|
Capabilities are indicated by three 32 bit flag values, and are
|
|
retrieved by the capabilities() function:
|
|
|
|
inst_mode cap;
|
|
inst2_capability cap2;
|
|
inst3_capability cap3;
|
|
|
|
it->capabilities(it, &cap, &cap2, &cap3);
|
|
|
|
The first set of flags correspond to the measurement mode capabilities.
|
|
|
|
Modes consist of basic mode combinations (light measurement mode + action),
|
|
+ dependent modes + independent modes.
|
|
|
|
It is generaly best to test and set for mode combinations rather than independent
|
|
flags, and to use the IMODETST(flags, mode) macro and check_mode() function
|
|
to test for a particular mode combination capability. (see the inst_mode enum).
|
|
|
|
For instance, if we are intending to do strip reflectance measurement, we
|
|
should test if it has this capability:
|
|
|
|
if (!IMODETST(cap, inst_mode_ref_strip)
|
|
|| it->check_mode(it, inst_mode_ref_strip) != inst_ok)
|
|
... error ...
|
|
|
|
Configuration consists of two aspects: setting the operating mode and setting options.
|
|
|
|
If we want to configure for reflective strip reading with spectral:
|
|
|
|
inst_mode mode = inst_mode_ref_strip;
|
|
|
|
if (!IMODETST(cap, inst_mode_spectral))
|
|
... error, no spectral support ...
|
|
|
|
mode |= inst_mode_spectral;
|
|
|
|
if ((rv = it->set_mode(it, mode)) != inst_ok)
|
|
... error ...
|
|
|
|
The exact capabilities may change with mode, so they should be re-fetched
|
|
if they are needed again:
|
|
|
|
it->capabilities(it, &cap, &cap2, &cap3);
|
|
|
|
For reflectance measurement with an X-Rite type instrument, we may want to
|
|
set the XRGA handling. i.e. to get xrga standard measurements:
|
|
|
|
if ((rv = it->get_set_opt(it, inst_opt_set_xcalstd, xcalstd_xrga)) != inst_ok)
|
|
... warning, XRGA not supported ...
|
|
|
|
Many instruments normally need a calibration every time they
|
|
are opened, but if several measurements are being done in quick
|
|
succession the calibration after opening the instrument
|
|
can be supressed by setting the inst_opt_noinitcalib option:
|
|
|
|
if ((rv = it->get_set_opt(it, inst_opt_noinitcalib, 0)) != inst_ok) {
|
|
... warning, not supported ...
|
|
|
|
6) Asynchronous callback handler
|
|
-----------------------------
|
|
|
|
To handle certain asynchronouos events, you can register
|
|
a callback handler. This function is typically called from
|
|
a different thread from the one making calls to inst functions.
|
|
|
|
Events are:
|
|
inst_event_switch: Instrument measure/calibrate/user trigger switch pressed.
|
|
inst_event_mconf: The measurement configuration has changed (ie. sensor position)
|
|
inst_event_scan_ready: Ready for manual strip scan (i.e. issue audible prompt to user)
|
|
|
|
Which events are issued will depend on the type of instrument and what mode and state
|
|
it is in.
|
|
|
|
void event_callback_handler(void *cntx, inst_event_type event) {
|
|
if (event == inst_event_scan_ready)
|
|
normal_beep();
|
|
}
|
|
|
|
it->set_event_callback(it, event_callback_handler, (void *)it);
|
|
|
|
If no event handler is registered, then the inst_event_scan_ready event will
|
|
generate a "beep" using operating system facilities.
|
|
|
|
7) Dealing with calibration
|
|
------------------------
|
|
|
|
Often it is better to do a proactive calibration, rather than waiting for
|
|
a calibration error to be returned by a measurement request. A need for
|
|
calibration can be determined by calling needs_calibration():
|
|
|
|
if (it->needs_calibration(it) & inst_calt_n_dfrble_mask)
|
|
... do needed calibration ...
|
|
|
|
Performing a calibration involves (usually) repeated calls to
|
|
calibrate(), with arguments *calt containing the mask of calibrations
|
|
to be performed, and *calc set to the current calibration condition.
|
|
|
|
Alternately, one of the psudo-calibration types inst_calt_all,
|
|
inst_calt_needed or inst_calt_available can be used,
|
|
and/or the *calc of inst_calc_none to get calibrate()
|
|
to determine the required calibration types and conditions.
|
|
(The corresponding calibration types will be used & returned.)
|
|
|
|
If no error is returned to the first call to calibrate(),
|
|
then the instrument was capable of calibrating without user or
|
|
application intervention. If on the other hand calibrate() returns
|
|
inst_cal_setup, then the appropriate action indicated by the value
|
|
returned in *calc should be taken by the user or application,
|
|
before retrying calibration() with the current setup indicated
|
|
by *calc. If more than one calibration type is requested, then
|
|
several retries may be needed with different calibration conditions.
|
|
Each call to calibrate() will update *calt to reflect the remaining
|
|
calibration to be performed. calibrate() returns inst_ok when no
|
|
more calibrations remain.
|
|
|
|
If the calc has the inst_calc_optional_flag flag set,
|
|
then the user should be offered the option of skipping the
|
|
calibration. If they decide to skip it, return a calc with
|
|
inst_calc_optional_flag set, and if they want to proceed,
|
|
make sure the inst_calc_optional_flag is cleared in the returned
|
|
calc.
|
|
|
|
The id parameter allows passing back messages to the user, such
|
|
as the serial number of the calibration plaque, or messages,
|
|
while the idtype parameter indicates the nature of what is
|
|
passed back in id, to allow for internationalization of
|
|
any messages.
|
|
|
|
The following is a cut down version of inst_handle_calibrate()
|
|
in the file instappsup.c.
|
|
|
|
inst_cal_type calt = inst_calt_needed;
|
|
inst_cal_cond calc = inst_calc_none;
|
|
inst_code ev;
|
|
int usermes = 0; /* User was given a message */
|
|
inst_calc_id_type idtype; /* Condition identifier type */
|
|
char id[200]; /* Condition identifier */
|
|
|
|
/* Untill we're done with the calibration */
|
|
for (;;) {
|
|
|
|
ev = p->calibrate(p, &calt, &calc, &idtype, id);
|
|
|
|
/* We're done */
|
|
if ((ev & inst_mask) == inst_ok) {
|
|
if ((calc & inst_calc_cond_mask) == inst_calc_message)
|
|
... display message id to user, or substitute our
|
|
own message based on the value of idtype ...
|
|
... done calibration ...
|
|
}
|
|
|
|
/* User aborted */
|
|
if ((ev & inst_mask) == inst_user_abort)
|
|
... done calibration with user abort ...
|
|
|
|
/* Retry on an error */
|
|
if ((ev & inst_mask) != inst_cal_setup) {
|
|
if ((ev & inst_mask) == inst_unsupported)
|
|
... requested calibration type is unsupported ...
|
|
|
|
... ask user whether to rety or abort,
|
|
and don't continue on abort ...
|
|
|
|
} else {
|
|
|
|
/* Ask user to do/setup calibration */
|
|
switch (calc & inst_calc_cond_mask) {
|
|
|
|
case inst_calc_man_ref_white:
|
|
... ask user to place instrument on its reflective white
|
|
reference S/N id string, and continue when done ...
|
|
break;
|
|
|
|
... cases not needed for i1pro have been omitted ...
|
|
|
|
default:
|
|
... we're not handling some calc ...
|
|
}
|
|
|
|
if (calc & inst_calc_optional_flag)
|
|
... offer user option to skip or abort calibration & measurement ...
|
|
else
|
|
... offer user option to abort calibration & measurement ...
|
|
|
|
... get user response ...
|
|
|
|
/* If optional calib. and user wants to skip it */
|
|
/* Loop back to calibrate() with inst_calc_optional_flag still set */
|
|
if ((calc & inst_calc_optional_flag) != 0 && ... user says skip ...)
|
|
goto oloop;
|
|
|
|
if (... user says abort ...)
|
|
... abort calibration & measurement ...
|
|
|
|
/* Remove any skip flag and continue with the calibration */
|
|
calc &= inst_calc_cond_mask;
|
|
}
|
|
oloop:;
|
|
}
|
|
|
|
8) Taking a single measurement
|
|
---------------------------
|
|
|
|
We may want to set a particular trigger mode before doing this. The appropriate
|
|
trigger mode depends on the instrument capabilities, as well as the way we
|
|
want to use it.
|
|
|
|
We can check what trigger modes are available by looking at the cap2 values
|
|
and comparing against the inst2_prog_trig, inst2_user_trig, inst2_switch_trig &
|
|
inst2_user_switch_trig flags, and then choose which mode to use. i.e. if we
|
|
want the user to initiate the measurement we might use:
|
|
|
|
/* Enable switch or user via uicallback trigger if possible */
|
|
if (cap2 & inst2_user_switch_trig) {
|
|
if ((rv = it->get_set_opt(it, inst_opt_trig_user_switch)) != inst_ok)
|
|
... error ...
|
|
|
|
/* Or go for user via uicallback trigger */
|
|
} else if (cap2 & inst2_user_trig) {
|
|
if ((rv = it->get_set_opt(it, inst_opt_trig_user)) != inst_ok)
|
|
... error ...
|
|
} else
|
|
... error, no usable trigger mode ...
|
|
|
|
or if we wanted to trigger progromatically:
|
|
|
|
if (cap2 & inst2_prog_trig) {
|
|
if ((rv = it->get_set_opt(it, inst_opt_trig_prog)) != inst_ok)
|
|
... error ...
|
|
|
|
} else
|
|
... error, no usable trigger mode ...
|
|
|
|
To measure one patch:
|
|
|
|
ipatch val;
|
|
|
|
rv = it->read_sample(it, "SPOT", &val, 1);
|
|
|
|
if (rv == inst_ok
|
|
|| (rv & inst_mask) == inst_user_trig)
|
|
... OK ...
|
|
|
|
else if ((rv & inst_mask) == inst_user_abort)
|
|
... abort measurements ...
|
|
|
|
else if ((rv & inst_mask) == inst_needs_cal)
|
|
... do calibration ...
|
|
|
|
else if ((rv & inst_mask) == inst_misread)
|
|
... deal with misread, i.e. retry ? ...
|
|
|
|
else
|
|
... some other error, probably fatal ...
|
|
|
|
ipatch values are in XYZ % (0..100) for reflective & transmissive,
|
|
cd/m^2 for emissive, and Lux for ambient.
|
|
Spectral will be analogous to the XYZ (see inst_meas_type).
|
|
By default values may be -ve due to noise (depending on instrument),
|
|
unless the clamp flag is set.
|
|
|
|
9) Taking strip measurements
|
|
-------------------------
|
|
|
|
We may want to set a particular trigger mode before doing this (see above).
|
|
|
|
Each strip may need certain parameters, depending on the type of instrument
|
|
used. For instance, a DTP51 displays a 7 character strip name,
|
|
and a 3 character pass name and a guide number, whereas an i1pro instrument
|
|
will ignore these parameters.
|
|
Similarly, the patch width, gap width and trailer width are only
|
|
needed by a DTP20 and DTP41.
|
|
|
|
To measure a strip:
|
|
|
|
#define NPATCHES 21
|
|
ipatch vals[NPATCHES];
|
|
|
|
rv = it->read_strip(it, "STRIP", NPATCHES, "001", 1, 7.366, 2.032, 18.8, vals);
|
|
|
|
if (rv == inst_ok
|
|
|| (rv & inst_mask) == inst_user_trig)
|
|
... OK ...
|
|
|
|
else if ((rv & inst_mask) == inst_user_abort)
|
|
... abort measurements ...
|
|
|
|
else if ((rv & inst_mask) == inst_needs_cal)
|
|
... do calibration ...
|
|
|
|
else if if ((rv & inst_mask) == inst_wrong_config)
|
|
... deal with bad sensor position, i.e. ColorMunki ...
|
|
|
|
else if ((rv & inst_mask) == inst_misread)
|
|
... deal with misread, i.e. retry ? ...
|
|
|
|
else
|
|
... some other error, probably fatal ...
|
|
|
|
10) Closing instrument
|
|
------------------
|
|
|
|
To close an instrument and the communication port, call the del() function:
|
|
|
|
|
|
it->del(it);
|
|
|
|
|
|
12) Error codes
|
|
-----------
|
|
|
|
The inst_code return codes generally consist of two parts; an instrument specific
|
|
16 bit error code in the lower 16 bits, and an abstract error code in the upper 8 bits.
|
|
|
|
These two parts can be separated using the inst_code inst_mask and inst_imask values.
|
|
|
|
An English language interpretation of the 16 bit instrument specific error code
|
|
can be obtained using the inst_interp_error() function:
|
|
|
|
it->inst_interp_error(it, rv)
|
|
|
|
and an English language interpretation of the 8 bit abstract error code
|
|
can be obtained using the interp_error() function:
|
|
|
|
it->interp_error(it, rv));
|
|
|