<h2> <u>Conversion to/from XRGA<br>
|
|
</u></h2>
|
|
After purchasing Gretag-Macbeth, X-Rite harmonized the subtly
|
|
different calibration standards the two companies were using for
|
|
their reflective measurement instruments, in order to minimize
|
|
changes with the introduction of new instruments. The result was the
|
|
<a href="http://www.xrite.com/xrite-graphic-arts-standard">XRGA
|
|
standard</a>, introduced in September 2010.<br>
|
|
<br>
|
|
Current X-Rite instruments are natively calibrated to XRGA, while
|
|
older instruments from X-Rite are calibrated to XRDI, and
|
|
Gretag-Macbeth instruments to GMDI standards. To smooth the
|
|
interchange of old and new instruments, a spectral <a
|
|
href="https://www.xrite.com/documents/literature/en/L7-462_XRGA_WhitePaper_en.pdf">conversion
|
|
|
|
|
|
|
|
|
|
|
|
|
|
is possible</a> between these different standards. While such a
|
|
conversion is not perfect, it reduces the discontinuities between
|
|
old instrument families and current X-Rite instruments.<br>
|
|
<br>
|
|
By default, Argyll will use the native calibration of Gretag-Macbeth
|
|
and X-Rite instruments. These are summarized here:<br>
|
|
<br>
|
|
<table cellspacing="2" cellpadding="2" border="0">
|
|
<tbody>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#DTP20">DTP20
|
|
"Pulse"</a> <br>
|
|
</td>
|
|
<td valign="top">XRDI<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#DTP22">DTP22
|
|
Digital Swatchbook</a></td>
|
|
<td valign="top">XRDI<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#DTP41">DTP41</a></td>
|
|
<td valign="top">XRDI<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#dtp51">DTP51</a></td>
|
|
<td valign="top">XRDI<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"> <a href="instruments.html#sl">Spectrolino</a>/
|
|
<a href="instruments.html#ss">SpectroScan</a></td>
|
|
<td valign="top">GMDI<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#i1p">Eye-One Pro</a>
|
|
Rev A-D</td>
|
|
<td valign="top">GMDI<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#ColorMunki"><span
|
|
style="text-decoration: underline;">ColorMunki</span></a>
|
|
Design, Photo or i1Studio<br>
|
|
</td>
|
|
<td valign="top">XRGA<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#i1p2">Eye-One Pro2</a>
|
|
(Eye-One Pro Rev E)</td>
|
|
<td valign="top">XRGA<br>
|
|
</td>
|
|
</tr>
|
|
<tr>
|
|
<td valign="top"><a href="instruments.html#i1p3">Eye-One Pro 3</a><br>
|
|
</td>
|
|
<td valign="top">XRGA<br>
|
|
</td>
|
|
</tr>
|
|
</tbody>
|
|
</table>
|
|
<br>
|
|
<br>
|
|
You can override this and force reflective data to be a particular
|
|
calibration standard in two ways:<br>
|
|
<br>
|
|
<a name="A"></a>Using the <b>-A</b> command line option in <a
|
|
href="spotread.html#A">spotread</a> and <a
|
|
href="chartread.html#A">chartread</a>:<br>
|
|
<br>
|
|
<b>-A </b>N|A|X|G<br>
|
|
<br>
|
|
The <b>N</b> argument sets
|
|
the calibration to Native.<br>
|
|
The <b>A</b> argument sets
|
|
the calibration to XRGA.<br>
|
|
The <b>X</b> argument sets
|
|
the calibration to XRDI.<br>
|
|
The <b>G</b> argument sets
|
|
the calibration to GMDI.<br>
|
|
<br>
|
|
<br>
|
|
Or by setting an environment variable:<br>
|
|
(Note that command line options will override this.)<br>
|
|
<br>
|
|
<br>
|
|
<span style="font-weight: bold;"><a name="ARGYLL_XRGA"></a>ARGYLL_XRGA</span><br>
|
|
<br>
|
|
<div style="margin-left: 40px;"> If the <span style="font-weight:
|
|
bold;">ARGYLL_XRGA </span> environment variable is set to:<br>
|
|
<br>
|
|
<b>XRGA</b>, then all Getag-MacBeth/X-Rite
|
|
reflective measurements will be converted to XRGA.<br>
|
|
<b> XRDI</b>, then all Getag-MacBeth/X-Rite
|
|
reflective measurements will be converted to XRDI.<br>
|
|
<b> GMDI</b>, then all Getag-MacBeth/X-Rite
|
|
reflective measurements will be converted to GMDI.<br>
|
|
<br>
|
|
If not set (or set to some other string), then the native
|
|
calibration will be used. Note that the values must be upper case.<br>
|
|
</div>
|
|
<span style="font-weight: bold;"></span><br>
|
|
All of the above except the <b>DTP51</b> can have a conversion
|
|
applied. <br>
|
|
(The DTP51 returns only colorimetric data, while the conversion
|
|
requires spectral data.)<br>
|
|
<br>
|
|
You can also convert from one standard to the other using <a
|
|
href="spec2cie.html#A">spec2cie</a>.<br>
|
|
<br>
|
|
For informational purposes, the native calibration of some other
|
|
Gretag-Macbeth and X-Rite instruments are:<br>
|
|
<br>
|
|
<table cellspacing="2" cellpadding="2" border="0">
|
|
<tbody>
|
|
<tr>
|
|
<td valign="top">HP Z printer Sensor <br>
|
|
</td>
|
|
<td valign="top">GMDI<br>
|
|
</td>
|
|
</tr>
|
|
</tbody>
|
|
</table>
|
|
<br>
|
|
<br>
|
|
<br>
|
|
</body>
|
|